AVE-733A
This urine analyzer operates on reflectance photometry and measures a wide range of parameters including blood, leucocytes, glucose, proteins, and kidney function indicators like microalbumin and creatinine. It offers high throughput of up to 514 tests per hour, with automatic error correction and self-diagnostics to ensure reliable results. The device features a touch screen interface, automatic strip handling, built-in and external printing options, and large storage capacity for over 2 million test results. Compact and lightweight, it maintains accuracy and stability under typical lab conditions while supporting both clinical and research needs.
Please Log In to access Additional Details and documents ..
Test Principle
Reflectance photometry
Detected Parameters
Occult blood, leucocytes, specific gravity, pH, protein, urobilinogen, bilirubin, ketones, glucose, nitrite, ascorbic acid, microalbumin, creatinine, calcium
Albumin/creatinine ratio (ACR) reportable
Features
High throughput (up to 514 tests/hour)
Automatic correction for non-specificity, pH, SG, and color interference
LCD touch screen, user-friendly interface
Automatic collection of used test strips
Built-in low-noise thermal printer, external printer support
Automatic saving/uploading of results
Power-on self-check, automatic fault reporting and error warning
Original NMPA-approved quality control materials available
Specifications
Throughput: ~514 tests/hour
Test strips: 10–14 parameters optional
Repeatability: CV ~1%
Accuracy: within one level of target values; no false negatives/positives with reference solutions
Stability: CV ~1% with 8 hours of continuous power
Storage: up to 2,100,000 results
Environmental: 30–80% relative humidity, 5–40 °C operating temp
Power: AC 100–240V, 50/60Hz, 250VA
Dimensions: 380 × 320 × 190 mmWeight: 4.5 kg